top of page
HiTemp140-FP Flexible Probe Temperature Data Logger

HiTemp140-FP Flexible Probe Temperature Data Logger

High Temperature Data Logger with Flexible RTD Probe. Includes ISO17025 accredited calibration certificate.

 

SPECIFICATIONS:

 

TEMPERATURE

Temperature SensorFlexible RTD Probe
Temperature Resolution0.01 °C (0.018 °F)

Temperature

Calibrated

Accuracy

±0.1 °C (0.18 °F)

Probe Temperature

Range

-60 °C to +260 °C (-76 °F to +500 °F) (Body of Logger Cannot Exceed 140 °C)

 

GENERAL

Memory32,767 readings
Reading Rate1 reading every 0.25 seconds up to 1 reading every 24 hours
Baud Rate125,000
Battery Life1 year typical (1 minute reading rate at 25 °C / 77 °F)
User Replaceable Battery
MaterialBody: 316 Stainless Steel
Probe: PFA Insulated Cable
DimensionsBody: 2.95 in x 0.97 in x 0.97 in (75 mm x 24.6 mm x 24.6 mm)
Probes: HiTemp140-FP-6: 6 in x 0.10 in (152.5 mm x 2.5 mm)
HiTemp140-FP-12: 12 in x 0.10 in (304.8 mm x 2.5 mm)
HiTemp140-FP-36: 36 in x 0.10 in (914 mm x 2.5 mm)
HiTemp140-FP-72: 72 in x 0.10 in (1829 mm x 2.5 mm)
IP RatingIP68
SubmersibleYes

Operating

Environment

-40 °C to +140 °C (-40 °F to +284 °F) 0 %RH to 100 %RH, 0 PSIA to 60 PSIA
With Thermal Shield: -200 °C to +250 °C (-328 °F to +482 °F) (Body of Logger Cannot Exceed 140 °C)

Required  Interface Package

IFC400 or IFC406
ApprovalsCE
Weight0.1875 LBS

 

The HiTemp140-FP is a durable, user friendly high temperature data logger featuring a long, flexible RTD probe with a narrow diameter, making it ideal for use in steam sterilization and lyophilization processes.

Commonly used for mapping, validation and monitoring of high temperature surfaces and environments, this stainless steel data logger is available in several flexible probe lengths. The flexible probe is coated with PFA insulation and can withstand temperatures up to 260 °C (500 °F) with an accuracy of ±0.1 °C.

The HiTemp140-FP probe design is narrow and lightweight making it ideal for placement within small vials, tubing, test tube and other small diameter or delicate applications. Because of the flexible probe, the risks of breakage (both vial and probe) generally associated with stainless steel probe loggers are diminished and the location and placement of the probe is easy to manipulate. The device records and stores up to 32,700 time stamped readings and is equipped with non-volatile solid state memory.

The HiTemp140-FP can be configured for delayed start and is capable of reading rates as often as four readings per second, up to once every 24 hours.

Compatible with the latest MadgeTech Data Logger Software, starting, stopping, and downloading data is simple and reporting capabilities are abundant. The device connects to a PC with the IFC400 interface or the IFC406 docking station (sold separately). Downloaded data can be viewed in graphical, tabular, and summary data form in the MadgeTech software, and all data can be exported to Excel® for further analysis and calculations.

Notice: Steam Sterilization Applications

The pervasive nature of pressurized steam creates a very difficult environment for electronics.
Please refer to the following preventative maintenance procedure when using this device in steam sterilization applications.

Additionally, this device is not ideal for steam sterilization applications above 121 °C/1.1 bar.