top of page
Cryo-Temp Data Logger

Cryo-Temp Data Logger

Ultra Low Temperature Data Logger.

Records Temperatures as low as -86 °C. Utilizes a USB Docking Station to communicate with a personal computer. (USB Docking Station sold separately). Includes ISO17025 accredited calibration certificate.

 

SPECIFICATIONS:

 

TEMPERATURE

Temperature SensorThermistor
Temperature Range-86 °C to +35 °C (-122.8 °F to +95 °F)
Temperature Resolution0.1 °C (0.18 °F)
Temperature Calibrated Accuracy±1.0 °C (1.8 °F)

 

GENERAL

Memory32,767
Reading Rate1 reading every 5 seconds up to 1 reading every 30 minutes
Single/Multi-UseMulti-use
Battery Life2 years typical at 25 °C, 1 min. reading intervals
1 year typical at -80 °C, 1 min. reading intervals
MaterialABS plastic
Dimensions3.3 in x 2.1 in x 0.4 in, (83 mm x 54 mm x 11 mm)
IP RatingIP64
Operating Environment-86 °C to +35 °C (-122.8 °F to +95 °F), 0 %RH to 95 %RH
Required Interface PackageIFC300
ApprovalsCE
Approved for carriage on Delta® Mainline Flights
Weight0.0875 LBS

New ultra low temperature data logger, records temperatures as low as -86 °C (-122.80 °F). The Cryo-Temp can be used in a broad line of applications such as monitoring blood plasma, vaccines, pharmaceuticals, frozen foods and shipping containers. This stand-alone device does not require any additional probes. The enclosure is designed with a handle for easy attachment and is splash proof. It utilizes a USB docking station to communicate with a personal computer. (USB docking station sold separately)

The Cryo-Temp provides date and time stamped temperature readings and uses the MadgeTech Data Logger Software. The user can view the data in graphical or tabular form. High and low alarm limits and high and low warning limits can be programmed through the software. The user can set the alarm limits in the range required for the goods to be maintained. The warn limits can be set to alert the user when the temperature is nearing a high or low alarm limit. The Cryo-Temp is equipped with three LEDs. The green LED will blink every 15 seconds while the recorder is logging. The yellow LED will blink every 3 seconds when the temperature has breached the warning limits. The red LED will blink every 3 seconds when the temperature alarm has been breached. The Cryo-Temp is the ideal temperature monitoring solution for any application involving low temperature data logging.)

Important Service Notice: To ensure optimum product performance and user safety, the Cryo-Temp data logger can not be serviced for calibration or battery replacement. As an alternative MadgeTech offers a product exchange program, please see the notification details here.